One-way
catheter valve.
Scroll to explore Inside the valve.
A hollow ball, a steel ball and a seal share one chamber.
Forward flow.
Urine passes around both balls toward the drainage bag.
Rising backflow.
The steel ball stays submerged. The hollow ball floats up to seal the inlet.
In any direction.
With the nozzle down, the steel ball presses the hollow ball against the seal. Urine fills the chamber up to the closed inlet.
Connection.
The nozzle connects to the Foley catheter. The drainage tube is integrated into the bottom of the valve.